|
Grounding and Shielding
Considerations for
Thermocouples, Strain Gauges, and Low-Level Circuits INTRODUCTION
Signal
to Noise Ratio
Electrical noise is present in all measurement systems to some
degree. Instrument engineers typically design systems that make
the measured signal much larger than the noise that accompanies
it, whatever that source of noise may be. They specify this parameter
as signal-to-noise ratio SNR and typically measure it in dB.
In terms of voltage, dB = 20 log (V2/V1), where V2 is greater
than V1, and V1 is the noise level. For example, 120 dB is a
ratio of 1,000,000:1 and 160 dB is 100,000,000:1, very respectable
numbers for SNRs. Design rules that instrument manufacturers
follow to make noise-free and quality equipment include proper
circuit design and layout on a printed circuit board as well
as grounding, shielding, and guarding. One common method for
measuring low-level signals where noise can be a problem is to
use differential input amplifiers. The topology of such circuits
tends to cancel certain types of noise. Both terminals of a two-terminal
sensor connect to the differential amplifier input, and shielding
and guarding circuits typically connect to signal ground and
special guard-shield terminals. This eliminates current flow
between the grounds, common, or returns of the sensor and the
instrument input amplifier, which can generate noise signals.
One drawback when using differential inputs is that only half
the number of sensor inputs is available compared to a single
ended configuration. That is, one differential amplifier consumes
two single-ended input connections.
DISCUSSION
High SNR is particularly critical for strain gauges and thermocouples,
which inherently have relatively low signal levels compared to
most sensors. These sensors deal with millivolts and microvolts,
signals already close to the values of typical noise levels that
can be measured in test setups. By comparison, many transducers
have 1V, 5V, and 10V standard output levels. Thus, 10V compared
to 1 microvolt is 10/0.000001, that is, 10 million to one or
20 log (10/0.000001) = 20 log (10,000000) = 20(7) = 140 dB. But
a 1.0 mV signal compared to a 1 microvolt level of noise is much
more serious. Here, 0.001V is compared to 0.000001V, or 20 log
(0.001/0.000001) = 20 log (1,000) = 20(3) = 60 dB. In other words,
the SNR for the first example is 10 million to one, and the second
is only 1 thousand to one. So, maintaining a high signal-to-noise
ratio is paramount in low-level signal circuits to prevent the
measured variable from being contaminated or yielding inaccurate
results.

PROCEDURE
Data acquisition system designers try to obtain the highest SNR
as possible within their equipment with the resources available
to them, but users also have an equal responsibility for reducing
noisy signals and maximizing instrument measurement accuracy.
Several such factors under user control that play a significant
role in reducing noise levels, increasing SNR, and improving
accuracy include transducer lead length, application of grounding
points and shields, and the temperature coefficient of copper
wire.

General Precautions
Short leads between sensors and amplifier inputs are the first
consideration in minimizing noise. Long leads act as antennas
and can pick up a variety of electric and magnetic interference.
Although strain gauges and thermocouples have characteristically
low impedances, it is still good practice to use short leads
and twist them together. Figure 1 illustrates a condition where
unshielded, parallel wires connect the signal source to the amplifier
input. The parallel wires pick up noise through induction from
the radiating wire principally because the mutual induction M1
and M2 are unequal as are the distances, d1 and d2.
Figure 2 shows how to cancel or minimize the interfering noise
by twisting the wires. Moreover, thermocouples should use thermocouple
lead wire rather than ordinary copper wire since TC wire is designed
just for this application.
In most cases, the special lead wire may be obtained from the
thermocouple manufacturer or supplier.
The second consideration is the ground point. Usually, a ground
connection is made at either the amplifier input or preferably
the sensor, but definitely not at both. When both devices are
grounded, a small but measurable "ground-loop" current
can flow between them and manifest in an unwanted noise signal
at the amplifier input terminals. The best policy is to follow
the manufacturers grounding recommendations for its particular
instrument and sensor.
The circuit in Figure 3 shows a potential difference between
earth ground #1 and earth ground #2. As a result, two ground
loop paths are generated when the shield, sensor return, and
sensor shield are all grounded to earth ground #1, and the other
end of the shield, the sensor return, the common amplifier terminal,
and amplifier chassis are connected to earth ground #2. One circulating
current runs through the shield conductor, and the other runs
through the signal return (or common) wire. Figure 4 shows that
the ground loop current through the shield is eliminated when
its ground connection is removed from earth ground #2. Likewise,
the loop current in the signal return path is eliminated when
the ground connection is removed from terminal 2 of the amplifier.
In addition, wires shielded with copper, aluminum, and tin can
minimize most electrical interference, but ferrous materials
may be needed to shield wires where high magnetic fields may
be present. Copper or aluminum shields alone may not be effective
enough. Moreover, grounding these shields require as much care
as other shields to minimize ground loop currents that can appear
as signals to the amplifier input terminals.

Copper wire
has a relatively high temperature coefficient of resistance.
However, it tends to be neglected when the sensor signals are
so large that any changes in impedance due to the wires
temperature coefficient have negligible effect. But in high impedance,
low-level circuits, or circuits with long lead wires, the wire
resistance by itself or with a change temperature can be significant
and affect the measurements by as much as 10% or more. In many
cases, however, wire resistance effects can be cancelled by using
a four wire system where one pair is the "excite" wires,
and the other pair is the "sense" wires, such as used
in bridge circuits.
Thermocouples

Thermocouples
should be electrically isolated from the device under test where
possible to avoid ground-loop currents, common-mode voltage problems,
and induced voltage or current. However, most thermocouple instrument
amplifiers provide ungrounded temperature reference input terminals
mounted on isothermal blocks that help minimize such noise problems.
For example, Figure 5 shows the temperature compensated isothermal
block and input connections to the IOtech DBK81 series
of thermocouple amplifiers.


When thermocouple
wire leads must be shielded, use the shield connection provided
at the input to the amplifier, and do not connect the other end
of the shield (at the device under test) to a ground. However,
when the thermocouple must be grounded, use the same ground terminal
to connect to the shield and do not use the shield connection
at the amplifier end. (Remember, the shield must not be grounded
at both ends.)
Strain gauges



Strain gauges (and frequently, thermocouples) are fastened mechanically
and sometimes electrically to the device under test. This arrangement
can set up a path for ground loop currents and common-mode voltage
noise signals. Use non-inductive types of strain gauges and ground
the shields only at the sensor end. Use the differential input
terminals of the data acquisition system and make certain that
neither terminal is inadvertently grounded through an inconspicuous
"sneak circuit." Figure 6 is an example of a situation
where a differential amplifier input is selected but the cable
shield is connected to the amplifier common. The common-mode
voltage Ecm appears between signal ground and system ground.
The distributed capacitance in the cable shield shunts the amplifier
input impedance to ground and sets up a noise current (Icm) path
through the signal source.
Removing the connection from the shield to system ground at the
amplifier input and connecting the shield to the signal ground
at the sensor as shown in Figure 7 reduces the effect of cable
capacitance and prevents common mode current, Icm, from flowing
through the signal source. In this way, the systems common mode
rejection ratio is greatly improved. Strain gauges and other
bridge circuits used with strain gauge modules such as the IOtech
WBK16 are not usually grounded. The WBK16 uses differential
input amplifiers with excitation voltage supplied by isolated
and current-limited power supplies to help minimize noise problems.
However, when strain gauge wires need shielding, use quality
cable, such as the CA-177 stain gauge cable, and connect the
shield to the DB9 metal shell. The shell does not return to common
ground, therefore common mode voltage and current cannot interfere
with the measured signal. Also, use twisted pair cable with paired
leads for signal input, excitation output, and remote sense input.
See Figure 8, the block functional diagram of the WBK16, for
an example showing connections to a full-bridge strain gauge.
Occasionally, some of the above precautions cannot be followed
implicitly. When in doubt, the best procedure is to use the differential
input terminals to the data acquisition system rather than single-ended
inputs, and connect the shield to the sensor or signal source
common terminal.

Other Considerations
The impedance of a signal source should be much lower than the
amplifier input impedance to ensure high system accuracy. As
the signal source impedance approaches zero, so does the amplitude
of the noise signal to the amplifier. However, in most cases,
users need not concern themselves with these issues. Most data
acquisition system manufacturers have considered such loading
problems and offer specific signal conditioning amplifiers that
match transducers and special circuits. For example, signal conditioners
are available for thermocouples and strain gauges that offer
the proper impedance matching and temperature compensation. However,
in signal conditioners that have more general capabilities such
as measuring voltage, current, and resistance, the sensors or
signal sources are not necessarily specified, and users should
consult the data acquisition system manufacturer to ensure that
they purchase the proper signal-conditioning amplifier. For example,
not all general-purpose voltage amplifiers are designed to work
with strain gauges, bridge circuits, and some other transducers
or circuits, especially those that could exceed the common-mode
voltage rating of the amplifier input. Its also a good
idea to discuss your test setup with an application engineer
to ensure that you will be properly connecting and grounding
your data acquisition system to the signal source, even when
you feel confident you have made the right decision.





Source Impedance
Calculations
For non-specified voltage and current measurements, the source
impedances must be known in order to select the appropriate signal
conditioner. Usually, the transducer manufacturer provides a
data sheet with the device that specifies its impedance and how
it is determined. When this is not available, a number of empirical
methods may be used to determine the source impedance. One method,
shown in Figure 9, is to measure the open circuit voltage output
with a known high impedance voltmeter. Next, connect a known
impedance to the same terminals and measure the voltage and current
under load. (Make certain that the impedance selected is not
so low as to cause the current to exceed the devices rating.)
The source (internal) impedance can be calculated form the equation:
Zi = Ze(V2V1)/V2
Where: Zi = internal impedance, Ohms
Ze = known external impedance, Ohms
V1 = measured open-circuit output voltage, V
V2 = measured closed-circuit output voltage, V
Typical
Environments and Tips
Electric motors, generators, and large transformers can present
unique problems. Large electromagnetic fields may be coupled
to sensor leads and induce noise. When measuring motor or generator
vibrations with accelerometers, notch filters may be used at
the amplifier input leads to eliminate 60 Hz signals that may
be induced. This is not a problem when the vibration frequencies
of interest are in the range of hundreds of Hz. In addition to
ac interference, dc motors and generators produce interfering
magnetic fields as well. Use the same shielding recommendations
as discussed above, using copper, aluminum, and ferrous materials
as appropriate for the environment.

Article kindly
provided by ICP Systems Ltd.
|

|